CONSTRUCTION AND CALIBRATION OF A CONDENSATION NUCLEUS COUNTER FOR THE MEASUREMENT OF AITKEN PARTICLES IN THE HARMATTAN HAZE

No Thumbnail Available
Date
1989-02-27
Authors
IBRAHIM, SHEHU MUSTAPHA
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
One type of atmospheric condensation nucleus i s the Aitken p a r t i c l e which has r a d i i ranging between 0.005 and 0.1 microns. In t h i s work, a locally constructed condensation nucleus counter s u i t a b l e for the measurement of the concentration of Aitken nuclei in the Harmattan haze is described. F i l t e r e d a i r is drawn into the previously evacuated . system to a pressure of about 2 x 104 Pa and then released. The concentration of the Aitken nuclei in the air drawn into the apparatus is deduced from the reading of the photocurrents Io and I through the pressurized and rarefied a ir in the system respectively. Measurements of p a r t i c l e s in ambient air with the counter show that the concentration of Aitken nuclei in the l o c a l i t y is up to 3 x 1011 p a r t i c l es per l i t r e.
Description
Submitted to the Postgraduate School, Ahnadu Bello University, Zaria, in Partial Fulfilment of the Requirements for 'the Degree of MASTER OF SCIENCE (NUCLEAR PHYSICS) DEPARTMENT OF PHYSICS AHMADU BELLO UNIVERSITY, ZARIA
Keywords
CONDENSATION NUCLEUS,, COUNTER,, MEASUREMENT,, AITKEN PARTICLES,, HARMATTAN HAZE
Citation