ASSESSMENT OF CHARACTERISTICS OF THERMOLUMINESCENCE DOSIMETRY (TLD) SYSTEM USED IN CENTRE FOR ENERGY RESEARCH AND TRAINING (CERT), AHMADU BELLO UNIVERSITY, ZARIA

No Thumbnail Available
Date
2008-11
Authors
BELLO, AISHA ADEMOH
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Thermoluminescence dosimetry (TLD) system has been used to estimate the personal dose for external occupationally exposed workers in CERT. The reliability and precision of dose measurements and the accuracy of dose evaluation are important factors for the improvement and achievement of individual monitoring objectives. In this piece of work, evaluation of characteristics of TLD system was investigated at the Centre for Energy Research and Training, Ahmadu Bello University Zaria. Dose sensitivity, dose linearity, residual signal, fading, batch homogeneity, stability of Calibration factor, re-usability over 10 cycles, sensitivity to ambient light, and the influence of environmental conditions were evaluated. Not all investigated properties were found to be within the limits defined by IEC standard. These results are discussed in order to demonstrate the degree of accuracy achieved and the need for its improvement where necessary.
Description
A THESIS SUBMITTED TO THE POST GRADUATE SCHOOL, AHMADU BELLO UNIVERSITY, ZARIA NIGERIA IN PARTIAL FULFILMENT FOR THE AWARD OF MASTER OF SCIENCE IN RADIATION BIOPHYSICS DEPARTMENT OF PHYSICS AHMADU BELLO UNIVERSITY, ZARIA NIGERIA NOVEMBER, 2008
Keywords
ASSESSMENT, CHARACTERISTICS, THERMOLUMINESCENCE DOSIMETRY (TLD) SYSTEM, CENTRE FOR ENERGY RESEARCH AND TRAINING (CERT, AHMADU BELLO UNIVERSITY,, ZARIA